cobas b 123 POC system 12 Troubleshooting
Status messages on the measurement report
Status messages on the measurement report
Status messages appear on measurement reports to provide you additional
information about your measurement result, or to explain why a measurement result
could not be determined.
Messages Explanation
(c) Correlation factors have been applied to this measurement result.
(q) This measurement result was determined by a parameter with a QC warning.
Base value not available Calculated value cannot be determined because an input value or measurement result is missing
or not available.
Hemolysis error Hemolysis cycle could not be successfully performed during the COOX measurement.
Improper sample type Measurement was attempted to be performed with an improper sample type.
Inhomogeneous sample Sample was not properly mixed before performing the measurement.
Insufficient volume Sample volume was not sufficient to measure the activated parameters.
Measurement in progress Measurement is still in progress and the result is not yet available.
Not calibrated Parameter was not calibrated before performing the measurement.
Operator cancel You have canceled the measurement before the results are available.
Spectral interference /
Interference error
Considerable percentage of the detected signal at the oximeter module comes from an interfering
substance. Results may be compromised.
Value above... Result is above the upper limit of the measuring range or the AMR.
Value below... Result is below the lower limit of the measuring range or the AMR.
Signal distorted Considerable percentage of the detected signal comes from an interfering substance.
Range error Calibration value is outside of the measuring range or the AMR.
Insufficient sample Air bubbles were aspirated while performing the measurement.
Slope error Result is outside of the measuring range or the AMR.
Linearity error Result is outside of the measuring range or the AMR.
Table 12-5 Status messages on the measurement report