EasyManua.ls Logo

Roche cobas b 123 - Page 231

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Instructions for Use · Version 13.0 231
cobas b 123 POC system 8 Calibration
Configuring calibration settings
p To define the daily frequency of 2P calibrations
q
To enable the automatic printing of calibration reports, the internal printer has to be
activated.
u To activate/deactivate the internal printer (p. 267)
1
Utilities > Configuration > Calibration > Times / intervals
2
To define the daily frequency of 2P calibrations, select 1 of the 3 interval options
on the 2 point calibration panel.
Figure 8-6 Calibration scheduler screen
The starting time of each 2P calibration is based on the starting time of the system
calibration.
For example, if you select the 8 h interval for 2P calibrations, and your start time
for system calibrations is 6:00 AM, 2P calibrations are performed on your
instrument at 2:00 PM, 10:00 PM and 6:00 AM.
s
p To activate/deactivate automatic printing of calibration reports
1
Utilities > Configuration > Calibration > Report > Auto report
Figure 8-7 Activate automatic report panel screen

Table of Contents

Related product manuals