Roche Diagnostics
Instructions for Use · Version 13.0 185
cobas b 123 POC system 7 Quality control
Performing QC measurements
2
In the AutoQC panel, press the button of one of the 3 QC levels that you want to
perform the QC measurement for.
The instrument performs the user-activated AutoQC measurement.
Figure 7-4 QC values tab during a user-activated AutoQC measurement
3
To enter the operator ID, press the button on the Operator ID field.
4
To accept or reject the results of the QC measurement, press the Accept or Reject
button.
Figure 7-5 QC values tab after a user-activated AutoQC measurement
s
p To perform a manual QC measurement
q
Define your QC materials before performing a manual QC measurement.
u Defining and configuring QC materials (p. 203)