7 Quality control cobas b 123 POC system
Performing QC measurements
q
COMBITROL PLUS B control material has to be acclimated to room temperature at least 24
hours before its use.
1
Workplace > QC measurement
The following window appears on the screen:
Figure 7-6 Quality control menu screen
2
To enter the QC material information, scan the barcode on the ampoule’s label.
Alternatively, you can manually enter this information by pressing
the button on the Manual QC panel and using the keyboard box that
appears on the screen. Then, press the OK button.
Use the QC material Insert Sheet to check the QC ranges entered.
3
Make sure that the QC material information that appears on the screen is correct.
Figure 7-7 QC measurement overview screen
4
To start the QC measurement, press the Start capillary measurement button.