Instructions for Use · Version 13.0 187
cobas b 123 POC system 7 Quality control
Performing QC measurements
5
Gently tap the head of the ampoule with your fingernail to remove any liquid
from the top.
Figure 7-8 Tapping the head of the ampoule
6
Break open the ampoule using gauze.
Figure 7-9 Opening the ampoule
CAUTION
Personal injury due to contact with glass splinters from broken QC ampoules
Glass splinters created when breaking QC ampoules may result in personal injury.
r When you break QC ampoules, use several layers of gauze and wear appropriate
personal protective equipment, such as:
o Lab gloves
o Lab coat
o Eye protection
o Face shield if there is a chance of splashing or splattering