Roche Diagnostics
188 Instructions for Use · Version 13.0
7 Quality control cobas b 123 POC system
Performing QC measurements
WARNING
Failure to follow QC protocols may lead to incorrect patient results
Failure to follow QC protocols may lead to incorrect patient results, which may
endanger patient lives.
r Follow quality control practices according to local regulations.
r Perform QC measurements within 30 seconds of opening the QC ampoule.
Immediately discard the ampoule after performing the manual QC measurement.
r Use an ampoule adapter to carry out the manual QC measurement.
7
Firmly attach the ampoule adapter to the fill port.
Figure 7-10 Attaching the ampoule adapater to the fill port
8
To aspirate the contents of the ampoule, press the Ye s button.
Figure 7-11 Aspirating the contents from the ampoule