Instructions for Use · Version 13.0 189
cobas b 123 POC system 7 Quality control
Performing QC measurements
9
When you receive the prompt on the screen to remove the capillary, remove the
ampoule adapter from the fill port and press Ye s .
Figure 7-12 Removing the ampoule adapter from the fill port
The instrument performs the manual QC measurement.
Figure 7-13 QC values tab during a manual QC measurement
10
To enter the operator ID, press the button on the Operator ID field.