Instructions for Use · Version 13.0 165
cobas b 123 POC system 6 Measurement
Performing a measurement
p To perform a measurement with a capillary or a
Roche MICROSAMPLER PROTECT
NOTICE
Using inappropriate capillary tubes may result in damage to the analyzer
Performing capillary measurements with inappropriate capillary tubes may damage the
analyzer.
r Roche recommends using its glass capillary tubes or
Roche MICROSAMPLER PROTECT for blood collection. If you use glass capillary tubes
from other manufacturers, ensure that they have heat-polished ends.
u Available accessories and consumables (p. 335)
1
Collect your sample for the measurement.
Measure your sample no later than 15 minutes after you have collected it.
2
In the Overview tab, select the parameters that you want to measure.
Parameters colored in dark green will be measured.
Parameters colored in light green will not be measured.
Figure 6-15 Overview tab
3
To start the measurement, press the Start capillary measurement.