EasyManuals Logo

Roche cobas b 123 User Manual

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #200 background imageLoading...
Page #200 background image
Roche Diagnostics
200 Instructions for Use · Version 13.0
7 Quality control cobas b 123 POC system
Configuring the QC scheduler
3
In the QC scheduler panel, press the New button.
The following window appears on the screen:
Figure 7-26 AutoQC time configuration screen
4
To set a start time, press the button in the Start time field. Enter your start
time using the numeric keyboard box that appears on the screen. Then, press the
OK button.
5
To select a QC level for the new time, press the button in the
AutoQC time setting dialog box that corresponds to the QC level that you want
to assign.
6
To apply your changes, press the OK button.
s
p To copy times in QC scheduler for one day to another day
1
Utilities > Configuration > Quality control > Times / intervals
Figure 7-27 QC scheduler screen
2
To select the day of the week that you want to copy times in QC scheduler for,
press the and buttons at the top of the QC scheduler panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 123 and is the answer not in the manual?

Roche cobas b 123 Specifications

General IconGeneral
BrandRoche
Modelcobas b 123
CategoryMedical Equipment
LanguageEnglish

Related product manuals