200 Instructions for Use · Version 13.0
7 Quality control cobas b 123 POC system
Configuring the QC scheduler
3
In the QC scheduler panel, press the New button.
The following window appears on the screen:
Figure 7-26 AutoQC time configuration screen
4
To set a start time, press the button in the Start time field. Enter your start
time using the numeric keyboard box that appears on the screen. Then, press the
OK button.
5
To select a QC level for the new time, press the button in the
AutoQC time setting dialog box that corresponds to the QC level that you want
to assign.
6
To apply your changes, press the OK button.
s
p To copy times in QC scheduler for one day to another day
1
Utilities > Configuration > Quality control > Times / intervals
Figure 7-27 QC scheduler screen
2
To select the day of the week that you want to copy times in QC scheduler for,
press the and buttons at the top of the QC scheduler panel.