Roche Diagnostics
208 Instructions for Use · Version 13.0
7 Quality control cobas b 123 POC system
Configuring QC rules and consequences
3
To define a QC rule for this parameter, press the and buttons in the Rules
panel.
The following QC rules are available:
4
To define a QC consequence for this parameter, press the corresponding button
in the Consequences panel.
To apply no QC consequences for this parameter, press the None button.
s
p To enable QC repeats for a parameter
q
Enabling the QC repeats feature allows the instrument to automatically repeat a QC
measurement at the same QC level if the rules for QC evaluation for selected parameters
are broken.
1
Utilities > Configuration > Quality control > Rules / conseq
Figure 7-39 Rules and consequences
2
To select a parameter to define QC rules and consequences for, press
the and buttons at the top of the QC rules and consequences panel.
3
To enable QC repeats for this parameter, make sure that the QC repeats button is
checked and highlighted. If the QC repeats button is not checked and
highlighted, press it.
s
None
No rule
RiliBÄK
RiliBÄK ranges
(1)
(1) Default RiliBÄK ranges for software versions 4.5 and lower are from the 2008
RiliBÄK release. Default RiliBÄK ranges for software versions 4.7 and higher are
from the 2014 RiliBÄK release.
Set 1
Rule 1 & Rule 2
Set 2
Rule 1, Rule 2 & Rule 4
Set 3
Rule 1, Rule 2, Rule 4 & Rule 5
Set 4
Rule 1, Rule 2, Rule 4, Rule 5 & Rule 6
2SD
Default