Improving Level Performance
R&S
®
SMA100B
213User Manual 1178.3834.02 ─ 05
Any R&S NRP sensor that supports the USB legacy protocol and is connected to one
of the USB interfaces, is detected automatically and added to the list. Vice versa, the
R&S SMA100B removes a sensor from the list, when it is disconnected.
R&S NRP sensors that are connected via LAN or use the USBTMC protocol are not
automatically detected. They are detected by the scan search function.
Access:
Select "Clk Syn/Power Sens" > "NRP Sensor Mapping".
The dialog lists all detected R&S NRP sensors for selection and mapping. You can
also browse the network for sensors.
The detected sensors are characterized by the used protocol and the correspond-
ing connector icon. In the "Mapping" column, you can assign the sensor to one of
the available sensor channels. The list can contain several entries but the
R&S SMA100B can only use up to four sensors simultaneously.
The remote commands required to define these settings are described in Chap-
ter 13.15, "SENSe, READ, INITiate and SLISt Subsystems", on page 498.
Settings
Sensor Mapping List .................................................................................................. 213
Scan ...........................................................................................................................214
Sensor Mapping List
Displays a list of all sensor entries with information on the sensor name, the used pro-
tocol, the connector and the assigned mapping.
If a sensor is connected via LAN or uses the USBTMC protocol, its protocol is indica-
ted as "Visa".
Using Power Sensors