General Instrument Functions
R&S
®
SMA100B
336User Manual 1178.3834.02 ─ 05
The "Disk & Memory" tab secures controlled access to the mass memory and pre-
vents information from leaving a secure environment.
The remote commands available to control security settings are described in Chap-
ter 13.17, "SYSTem Subsystem", on page 653.
USB Storage
Activates the access to external USB storage media.
To apply the change: enter the security password and confirm with "Accept". Otherwise
the change has no effect.
See also Chapter 10.9.4, "Using a USB Storage Device for File Transfer",
on page 310.
Volatile Mode
Activates volatile mode, so that no user data can be written on the internal memory
permanently.
In volatile mode:
Data that the instrument normally stores on the internal memory is redirected to
volatile memory.
The user directory is mapped to the volatile memory. You access the temporary
data just as data stored in the /var/user/, see Chapter 10.3, "Protecting Data",
on page 290.
Data on the internal memory cannot be changed. It is protected against modifica-
tion or erasure.
You can only save data:
Temporarily in the volatile memory
On a connected external storage device, such as a memory stick
To activate volatile mode: enter the security password, confirm with "Accept" and
reboot the instrument. Otherwise the change has no effect.
Activated volatile mode is indicated by an icon.
Remote command:
:SYSTem:SECurity:VOLMode[:STATe] on page 662
Sanitize
Executes the erase procedure that sanitizes the internal memory.