Remote Control Commands
R&S
®
SMA100B
464User Manual 1178.3834.02 ─ 05
Manual operation:  See " Auto Set - Power Analysis " on page 250
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:DTIMe <DTime>
Determines the minimum time for which the signal must be below (above) the power
level defined by level and hysteresis before triggering can occur again.
Parameters:
<DTime> float
Range:  0  to  10
*RST:  200E-9
Example: 
SENS1:SWE:TIME:SENS:TRIG:DTIM 10 us
The drop out time is 10 us
Manual operation:  See " Drop out Time - Power Analysis " on page 250
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:HYSTeresis 
<Hysteresis>
Sets the hysteresis of the internal trigger threshold. Hysteresis is the magnitude (in dB)
the trigger signal level must drop below the trigger threshold (positive trigger slope)
before triggering can occur again.
Parameters:
<Hysteresis> float
Range:  0  to  10
Increment:  0.001
*RST:  0.5
Example: 
SENS1:SWE:TIME:SENS:TRIG:HYST 0.5 dB
The hysteresis is 0.5 dB
Manual operation:  See " Hysteresis - Power Analysis " on page 249
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:LEVel <Level>
Sets the trigger threshold.
Parameters:
<Level> float
Range:  -200  to  100
Increment:  0.001
*RST:  1
Example: 
SENS1:SWE:TIME:SENS:TRIG:LEV -20 dBm
Sets the trigger level to -20 dBm.
Manual operation:  See " Level - Power Analysis " on page 249
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:SLOPe <TriggerSlope>
Sets the polarity of the active slope for the trigger signals.
Power Sensor Measurement Subsystems