Remote Control Commands
R&S
®
SMA100B
463User Manual 1178.3834.02 ─ 05
Parameters:
<Reference> float
Range: 0.0 to 100.0
Increment: 0.01
*RST: 50.0
Example:
SENS1:SWE:TIM:PULS:THR:REF 40
Sets the medial reference level to 40% of the overall pulse level.
Options: Option R&S SMAB-K28
Manual operation: See " Mesial - Pulse Data Analysis " on page 244
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:SFRequency
<SFrequency>
Defines the separate frequency used for power vs. time measurement.
Parameters:
<SFrequency> float
Range: 0 to 1E12
Increment: 1
*RST: 1E6
Example:
SENS1:SWE:TIME:SENS:SFR 2GHz
The measurement is performed at 2 GHz
Manual operation: See " Use Separate Frequency- Power Analysis " on page 240
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:SFRequency:STATe <State>
Activates the use of a different frequency for the power measurement.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:SWE:TIME:SENS:SFR:STAT ON
Activates the use of a separate frequency than the generator fre-
quency for power analysis
Manual operation: See " Use Separate Frequency- Power Analysis " on page 240
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:AUTO <Auto>
Sets the trigger level, the hysteresis and the dropout time to default values.
Parameters:
<Auto> ONCE
*RST: ---
Example:
SENS1:SWE:TIME:SENS:TRIG:AUTO ONCE
The trigger level is automatically determined
Power Sensor Measurement Subsystems