Remote Control Commands
R&S
®
SMA100B
462User Manual 1178.3834.02 ─ 05
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
HREFerence <HReference>
Sets the upper reference level in terms of percentage of the overall pulse level (power
or voltage). The distal power defines the end of the rising edge and the start of the fall-
ing edge of the pulse.
Note: The command is only available in time measurement mode and with R&S
NRPZ81 power sensors.
Parameters:
<HReference> float
Range: 0 to 100
Increment: 0.01
*RST: 90
Options: Option R&S SMAB-K28
Manual operation: See " Distal - Pulse Data Analysis " on page 244
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
LREFerence <LReference>
Sets the lower reference level in terms of percentage of the overall pulse level. The
proximal power defines the start of the rising edge and the end of the falling edge of
the pulse.
Note: This parameter is only available in time measurement mode and R&S NRP-Z81
power sensors.
Parameters:
<LReference> float
Range: 0.0 to 100.0
Increment: 0.01
*RST: 10.0
Example:
SENS:SWE:TIM:PULS:THR:LREF 10
Sets the lower reference level to 10%.
Options: Option R&S SMAB-K28
Manual operation: See " Proximal - Pulse Data Analysis " on page 244
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
REFerence <Reference>
Sets the medial reference level in terms of percentage of the overall pulse level (power
or voltage related). This level is used to define pulse width and pulse period.
Note: The command is only available in time measurement mode and with R&S
NRPZ81 power sensors.
Power Sensor Measurement Subsystems