Remote Control Commands
R&S
®
SMA100B
470User Manual 1178.3834.02 ─ 05
Parameters:
<Minimum> float
Range: -200 to 100
Increment: 0.01
*RST: -40
Example:
SENS:SWE:FREQ:YSC:MIN -10DB
Sets -10 dBm as the lower limit of the measurement diagram.
Manual operation: See " Min - Max Level - Power Analysis " on page 235
:SENSe[:POWer]:SWEep:HCOPy:DATA?
Queries the measurement data directly. The data is transferred to the remote client as
data stream.
Readable ASCII data is available for hardcopy language CSV. The representation of
the values depends on the selected orientation for the CSV format.
Return values:
<Data> block data
Example:
SENS:SWE:HCOP:DEV:LANG CSV
selects output format *.csv.
SENS:SWE:HCOP:DEV:LANG:CSV:ORI HOR
selects horizontal orientation
SENS:SWE:HCOP:DEV:LANG:CSV:SEP SEM
selects ";" as the separator between the values
SENS:SWE:HCOP:DEV:LANG:CSV:DPO DOT
selects "." as decimal point
SENS:SWE:HCOP:DATA?
queries the measurement data of the current traces
Response:
#2651009500000;1019000000;1028500000;1038000000
-9.5;-9.7;-6.3;-2.5
The hash symbol # introduces the data block. The next number
indicates how many of the following digits describe the length of
the data block. In the example, the 2 following digit indicates the
length to be 65 characters.
Because horizontal representation is selected, a row with all the
x-values of the active trace (frequency) follows. The second row
contains all the y-values of the active trace (power). The rows
end with a new line (each counts as one character).
Note: if more than one trace is active, the third row contains the
x values of the second active trace, and so on.
Power Sensor Measurement Subsystems