Manual operation: See "Prefix, Year, Month, Day, Instrument Name" on page 257
:SENSe[:POWer]:SWEep:HCOPy[:EXECute]
Triggers the generation of a hardcopy of the current measurement diagram. The data
is written into the file selected/created with the :SENSe[:POWer]:SWEep:HCOPy:
FILE[:NAME] on page 473 command.
Example:
SENS:SWE:HCOP:DEV:LANG BMP
Selects output format *.bmp.
SENS:SWE:HCOP:FILE:AUTO:STAT OFF
Switches off automatic file naming.
SENS:SWE:HCOP:FILE 'var/nrp_trace1'
Creates the file nrp_trace1.bmp in the set path.
SENS:SWE:HCOP
Triggers the generation of a hardcopy of the current measure-
ment diagram. The hardcopy is stored in the file
nrp_trace1.bmp.
Usage: Event
Manual operation: See " File name - Power Analysis " on page 254
:SENSe[:POWer]:SWEep:INITiate
Starts the power analysis with NRP power sensor.
Example:
SENS:SWE:INIT
Start the power measurement.
Usage: Setting only
:SENSe[:POWer]:SWEep:MODE <Mode>
Selects power versus frequency measurement (frequency response), power vs power
measurement (power sweep, AM/AM) or power vs. time measurement.
Parameters:
<Mode> FREQuency | POWer | TIME
*RST: FREQuency
Example:
SENS:SWE:MODE FREQ
Selects frequency mode.
Manual operation: See " Mode - Power Analysis " on page 231
:SENSe[:POWer]:SWEep:POWer:REFerence:DATA:COPY
Generates a reference curve for "Power" measurement.
Example:
SENS:POW:SWE:POW:REF:DATA:COPY
Generates a reference curve in power mode.
Power Sensor Measurement Subsystems