Remote Control Commands
R&S
®
SMA100B
632User Manual 1178.3834.02 ─ 05
Parameters:
<Amplitude> float
Range: depends on the installed options
Increment: 0.01
*RST: n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation: See " Limit " on page 69
[:SOURce<hw>]:POWer:LMODe <LevMode>
Sets the RF level mode.
Parameters:
<LevMode> NORMal | LOWNoise | LOWDistortion
NORMal
Supplies the RF signal with the standard power level of the
instrument.
LOWNoise
Supplies a very low noise sinewave signal.
LOWDistortion
Supplies a very pure sinewave signal.
*RST: NORMal
Example:
SOURce1:POWer:LMODe LOWD
Sets low distortion mode. The instrument reduces distortions of
the RF signal to a minimum.
Manual operation: See " Mode " on page 70
[:SOURce<hw>]:POWer:MANual <Manual>
Sets the level for the subsequent sweep step if SWE:POW:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any level within the setting range, where:
STARt is set with [:SOURce<hw>]:POWer:STARt
STOP is set with [:SOURce<hw>]:POWer:STOP
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
Example: See Example "Setup an RF frequency or power sweep"
on page 643
SOURce Subsystem