Remote Control Commands
R&S
®
SMA100B
646User Manual 1178.3834.02 ─ 05
Parameters:
<Dwell> float
Range: 3E-3 to 100
Increment: 100E-6
*RST: 10E-3
Default unit: s
Example: See Example "Setup an RF frequency or power sweep"
on page 643.
Manual operation: See " Dwell Time " on page 174
[:SOURce<hw>]:SWEep:GENeration <SweepType>
Selects frequency sweep type.
Parameters:
<SweepType> STEPped | ANALog
STEPped
Performs a frequency sweep.
ANALog
Performs a continuous analog frequency sweep (ramp),
synchronized with the sweep time [:SOURce<hw>]:SWEep[:
FREQuency]:TIME.
*RST: STEPped
Manual operation: See " Sweep Type " on page 171
[:SOURce<hw>]:SWEep:POWer:MODE <Mode>
Sets the cycle mode for the level sweep.
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually with the command [:SOURce<hw>]:POWer:MANual.
The level value increases at each step by the value that you
define with [:SOURce<hw>]:POWer:STEP[:INCRement].
Values directly entered with the command [:SOURce<hw>]:
POWer:MANual are not taken into account.
STEP
Each trigger triggers one sweep step only. The level increases
by the value entered with [:SOURce<hw>]:POWer:STEP[:
INCRement].
*RST:
AUTO
SOURce Subsystem