5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
17. Press the "Down"-Key.
The BASKET lid will open completely.
18. Remove the BASKET lid from the holder on the DAC UNIVERSAL at
the end of the cycle to allow the instruments to cool down faster.
19. Place the BASKET lid on the counter or put it in the optional lid holder.
20. Check the instruments visually to make sure they are dry.
21. Check the instruments visually to make sure they have been cleaned.
22. Check the cycle number, properly conducted cycle on your
documentation (printer, software, etc.): For sterilization, "Sterilization,
time, ok" must be indicated.
23. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
24. Document the cycle according to the work instructions applicable in
your practice.
25. Open the snap lock on the basket and take the small lid off the basket.
26. Remove the solid instruments from the BASKET lid.
27. Place the small lid back on the basket.
The solid instruments can now be used again.
Sterilization of unwrapped instruments is not sufficient for invasive
procedures. After unwrapping the instruments another sterilization
WARNING
After completion of the cycle, make sure that all solid instruments are
still in the basket. If instruments have fallen into the chamber during the
cycle, cleaning and sterilization have failed. In this case, reprocessing of
the instruments must be repeated.
CAUTION
Danger of burning!
Even when the BASKET lid is half or completely open, the instruments
are still hot. Allow the extremely hot instruments to cool down after the
process.
CAUTION
Danger of burning!
The metal part of the lid is very hot. The plastic part of the basket can be
hot. Please touch only the plastic part of the BASKET lid at the end of
the cycle.