60 53 362 D3472
166 D3472.201.10.20.02 07.2016
8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.2Using the menu and display information DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
8.2.2.2
Menu overview DAC UNIVERSAL STANDARD
Menu over view with FLEX lid
Menu level 1 Menu level 2 Menu level 3 Result / Activity / Note
"Program" "134 °C"
Program: 134°C / 273°F, Sterilization time 3 min.
"121 °C"
Program: 121°C / 250°F, Sterilization time 15 min.
"Ready 134 °C
18min"
Program: 134°C / 273°F, Sterilization time 18 min.
"Manual" "Open"
Opens the lid.
"Close"
Closes the lid.
The lid must be in place.
"Water"
Pumps water into the chamber.
"Draintank"
Empties the water tank through the "Water" inlet at the back
of the unit. Automatic water supply must be interrupted.
"Draincham"
Empties the chamber theough the "Drain" connection.
The lid must be in place.
The lid is raised approx. 60 seconds after starting this function
and pressing the "Menu/Enter"-Key.
The chamber should then be empty. If not, the operation can
be repeated.
"Flow test"
Measurement of inlet air flow.
"Settings" "Printer" "ON "
/
"OFF"
Switches the printer on or off.
"Printout
language"
Selects the language of the print. The available printout
languages are English, German and French.
"Reprint"
The last print is printed again.
"Rinse cycles"
Rinse cycles (4-
6)
Standard setting: 4 cleaning cycles. (3 cold, 1warm) Can be
increased to up to 6 cleaning cycles (5 cold, 1 warm). The
FLEX lid will always use 6 rinse cycles (5 cold,1 warm).
Please note that the total cycle time is prolonged with more
than 4 cleaning cycles and water consumption increases.
"Oil expel"
1 - 5
Oil purge
Condensation and excess oil are purged from the instruments
with air.
Standard setting: Level 1. The maximum is level 5.
Please note that oil is purged before sterilization and that
increasing oil purging prolongs the total cycle time.
Level 1 is recommended for instruments of the manufacturers
W+H, KaVo, NSK, and most other instruments. Level 3 is
recommended for Sirona T1 CLASSIC/TE. However, please
observe the manufacturer's instructions for use for each
instrument.
"Language"
UK English is selected as display language.
D German is selected as display language.
F French is selected as display language.
SP Spanish is selected as display language.
I Italian is selected as display language.
S Swedish is selected as display language.