6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S
6.10Cleaning the empty chamber of the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
Process
Checking adapters with FLEX lid
✔ The lid is at hand and cooled down.
1. Place the lid on the counter or put it in the optional lid holder.
2. Check the attachment of the adapters to the lid by pulling each
adapter lightly.
3. If an adapter is loose, tighten the screws with the 1.5 mm Allen
wrench included with the unit.
4. Repeat for all available STANDARD or FLEX lids.
6.10
Cleaning the empty chamber of the DAC
UNIVERSAL ADVANCED
Frequency: weekly or every 50 cycles (whichever comes first).
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ The STANDARD lid of the DAC UNIVERSAL is removed and there is
an unobstructed view into the chamber.
✔ There is no water or oil in the chamber or the hoses.
1. Clean the empty chamber with a lint-free cloth and a solution of 70%
isopropyl alcohol alcohol.
2. Insert the STANDARD lid in the cover holder of the DAC
UNIVERSAL.
IMPORTANT
If an O-ring on the adapters is damaged or missing, replace it, see
"Changing the O-rings on the adapters [ → 139]".
IMPORTANT
The chamber may not be cleaned with agents containing chlorine.