8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
General attachment of scaling tips
1. The scaler tip was unscrewed from the ultrasonic handpiece with the
torque wrench and the tip remains in the torque wrench.
2. Ensure that the torque wrench is correctly attached to the tip
General placement
The instruments should not be leaning on each other; otherwise they can
be attached as preferred by the user.
Placement is described in the separate sections if it deviates from the
general placement described here.
See "Adapters for the FLEX lid [ → 238]" for detailed instructions on the
individual adapters.
8.3.5.2
Start cycle on the FLEX lid DAC UNIVERSAL STANDARD
✔ All settings of the DAC UNIVERSAL were preset according to
"Regular activities [ → 153]".
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ The maximum number of 6 instruments are ready for reprocessing.
✔ The FLEX lid is at hand.
1. Remove the FLEX lid from the holder on the DAC UNIVERSAL.
2. Place the FLEX lid on the counter or put it in the optional lid holder.
3. Plug each instrument onto the corresponding adapter in the FLEX lid.
See "General placement of instruments DAC UNIVERSAL
STANDARD [ → 187]" for more information on attaching the
instruments. The adapters can be replaced if necessary, see
"Mounting adapters [ → 53]".
4. Attach a suitable chemical indicator in the FLEX indicator holder in
the groove on the center post of the FLEX lid. See "Checking the
reprocessing result with an indicator / PCD test device [ → 153]" for
more information on using the chemical indicator / PCD test device.
WARNING
Scaler tips must be inserted with the associated torque wrench. The
torque wrench should remain on the tip during the whole process.
NOTICE
Tips have to be screwed straight onto the adapter so they can’t jam up.
WARNING
The instruments may not be wrapped when using the FLEX lid, as
cleaning and disinfection would then fail.
IMPORTANT
Do not start the DAC UNIVERSAL without a lid inserted into the lid
holder.