EasyManua.ls Logo

Sirona DAC UNIVERSAL - Cleaning the drain filter

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
130 D3472.201.10.20.02 07.2016
6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S
6.7Cleaning the drain filter Operating Instructions DAC UNIVERSAL
9. Wipe the sieve dry with a clean cloth.
10. Grasp the recesses of water tank with both hands again.
11. Insert the water tank back into the DAC UNIVERSAL.
12. Wriggle slightly until the water tank is firmly in place.
13. Place the sieve back on the water tank.
14. For manual water supply: Fill demineralized or distilled water (water
conductivity < 3.0 µS/cm) into the water tank.
15. Reconnect the DAC UNIVERSAL to the power supply.
The DAC UNIVERSAL is now switched on and ready for operation
again.
6.7
Cleaning the drain filter
Frequency: weekly or every 50 cycles (whichever comes first).
The DAC UNIVERSAL is switched off and cooled down.
You are wearing your protective gear.
The filter key included with the unit is at hand.
1. Unscrew the lid of the filter housing using the filter key.
NOTICE
When inserting the water tank, make sure the float is put through the
opening in the water tank.
IMPORTANT
At approx. 10 cycles per day, the drain filter must be changed every 4
weeks, see "Changing the drain filter [ 138]"

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals