9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S
9.6Cleaning the water tank Operating Instructions DAC UNIVERSAL
9. Press the "Menu/Enter"-Key to confirm the selection.
10. Press the "Up" or "Down"-Key until
"Draintank"
is selected.
11. Press the "Menu/Enter"-Key to confirm the selection.
The water is drained from the tank through the hose into the
drain/sink or into a bucket.
This process takes approx. 5-10 minutes.
12. Press the "Menu/Enter"-Key when thewater tank is completely
empty.
13. Press the "Clear"-Key until the normal display reappears.
The tank is empty.
14. Open the "Water" quick coupling by twisting it counterclockwise.
15. Detach the hose used for draining the water tank.
16. Reattach the normal 6/4-mm hose to the "Water" quick coupling at
the back of the unit and twist it clockwise firmly.
9.6
Cleaning the water tank
Frequency: weekly or every 50 cycles (whichever comes first).
✔ The water tank is empty, see "Drain the water tank of the DAC
UNIVERSAL STANDARD [ → 196]".
✔ The DAC UNIVERSAL is switched off.
1. Remove the sieve from the water tank
IMPORTANT
Step 16 is required only if there is a permanent water connection.
IMPORTANT
If you have installed the DAC UNIVERSAL with a permanent water
connection, please remember to turn the water back on.
IMPORTANT
If there is no permanent water connection, fill the water tank with
demineralized or distilled water (water conductivity < 3.0 µS/cm).