7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S
7.1Display messages and what they mean DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
7.1
Display messages and what they mean DAC
UNIVERSAL ADVANCED
The display messages each have an assigned number. The number will
be visible in the top left corner, next to the error symbol (Not necessarily
shown below). The unique code number for each display message is
shown in the table below, combined with an explanation to the specific
display message:
Code
no.
Pictogram Description Corrective action
60 The chamber
temperature and the
theoretical steam
temperature deviate from
each other more than 2K.
Please contact your authorized dealer.
61 The software has not
registered a change in
temperature in the steam
generator for 3 minutes.
Please contact your authorized dealer.
62 The temperature required
for building up pressure
for sterilization was not
reached within 7½
minutes.
Increase the level of the "Ambient temperature
sensitivity" settings.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC UNIVERSAL is in a room with a
temperature < 16°C / 61°F, ensure that operating
conditions is fullfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC UNIVERSAL.
If the error occurs repeatedly, please contact your
authorized dealer.
63 Incorrect setting of the
water level switch.
Please contact your authorized dealer.