60 53 362 D3472
136 D3472.201.10.20.02 07.2016
6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S
6.11Cleaning the housing of the unit Operating Instructions DAC UNIVERSAL
11. Remove the STANDARD lid from the holder on the DAC UNIVERSAL
at the end of the cycle.
12. Place the STANDARD lid on the counter or put it in the optional lid
holder.
The chamber and hoses are clean.
6.11
Cleaning the housing of the unit
Frequency: weekly or every 50 cycles (whichever comes first).
The DAC UNIVERSAL is switched off and cooled down.
The DAC UNIVERSAL is switched off and cooled down.
6.12
Cleaning the lids
Frequency: weekly or every 50 cycles (whichever comes first).
CAUTION
Danger of burning!
The metal part of the lid is very hot. Please touch only the plastic part of
the lid at the end of the cycle.
IMPORTANT
The surface may not be cleaned with agents containing chlorine.
CAUTION
The metal part of the lid is very hot at the end of the cycle. Please allow
the lid to cool down before cleaning it.