9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S
9.8Remove and clean the drain tank Operating Instructions DAC UNIVERSAL
9.8
Remove and clean the drain tank
Frequency:
If using manual water: Weekly or every 50 cycles (whichever comes first)
or when it is full.
If using automatic water: Daily or every 50 cycles (whichever comes first)
or when it is full.
✔ The DAC UNIVERSAL is switched off and cooled down.
✔ The waste water tank is cooled down.
✔ You are wearing your protective gear.
1. Unscrew the lid of the drain tank.
NOTICE
Do not start a cycle if the DAC UNIVERSAL is not connected to the drain
tank or drainage system.