EasyManua.ls Logo

Sirona DAC UNIVERSAL - Checking oil supply to the chamber of the DAC UNIVERSAL ADVANCED; Checking water supply to the chamber of the DAC UNIVERSAL ADVANCED

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
122 D3472.201.10.20.02 07.2016
6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S
6.1Checking oil supply to the chamber of the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
6.1
Checking oil supply to the chamber of the DAC
UNIVERSAL ADVANCED
Frequency: weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC UNIVERSAL is able to lubricate the
instruments correctly, the oil flow to the chamber needs to be tested.
The DAC UNIVERSAL is switched on and ready for operation.
The NitramOil #2 bottle is properly screwed into the DAC
UNIVERSAL.
The lid of the DAC UNIVERSAL is removed and there is an
unobstructed view into the chamber.
1. Enter the "manual oil operation“ menu section, by pressing the "Left“-
Key.
2. Press the "Down"-Key.
The oil dispenser will start to pump and oil flows through the
bottom adapter into the chamber.
3. Repress the "Down“-Key when oil is visible.
The display of the DAC UNIVERSAL counts the number of times the
oil dispenser pumps.
The water and oil residue that is in the chamber and hoses will need
to be drained and cleaned out, see section "Drain the chamber" in
"Drain the water tank via the menu [ 125]".
6.2
Checking water supply to the chamber of the DAC
UNIVERSAL ADVANCED
Frequency: weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC UNIVERSAL is able to bring water to
the chamber correctly, this operation needs to be tested.
The DAC UNIVERSAL is switched on and ready for operation.
The lid of the DAC UNIVERSAL is removed and there is an
unobstructed view into the chamber.
1. Enter the "manual water operation“ menu section.
2. Press the "Down"-Key.
IMPORTANT
If the number indicated is >60 and there is no oil in the chamber, please
contact your authorized dealer.

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals