60 53 362 D3472
54 D3472.201.10.20.02 07.2016
4Installation Sirona Dental A/S
4.4Mounting adapters Operating Instructions DAC UNIVERSAL
4.4.2 Mounting adapters on the FLEX lid
The FLEX lid can contain 6 instruments in total, in 4 different
configurations. These are following:
3 ultrasonic scaler handpieces or sprayvit nozzles and 3 scaler tips
w/ torque wrench at a time.
4 ultrasonic scaler handpieces or sprayvit nozzles and 2 scaler tips
w/ torque wrench at a time.
5 ultrasonic scaler handpieces or sprayvit nozzles and 1 scaler tips
w/ torque wrench at a time.
6 ultrasonic scaler handpieces or sprayvit nozzles.
The adapters are setup according to preferences and instruments in the
practice.
Fastening the adapter to the FLEX lid
The FLEX lid and the 6 adapters are available.
1. Place the O-rings on the FLEX lid in there meant fitting at each of the
6 adapter locations.
2. Place the adapter over the O-ring. Make sure that the flat side of the
adapter base is turned towards the center column as shown.
IMPORTANT
When mounting the adapters it is important that all O-rings, gaskets and
screws are new.