5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
5.2.3 Standard settings and Resetting the menu DAC
UNIVERSAL ADVANCED
Standard settings
Only for Japan: ambient temperature sensitivity is set at 4. Please note
this when resetting the menu to standard settings the ambient
temperature sensitivity will be set to 1.
Resetting the menu system
To reset the menu to the default settings, please proceed as follows:
The display shows the ready screen.
Press the "Down"-Key and the "Left"-Key simultaneously for more
than two seconds.
The properties in your menu are reset to the default settings.
Program 134°C / 273°F, Sterilization time 3 min.
Printer: Off
Language: UK
Oil expel: 1
Oil settings: 1
Rinse cycles: 4
Expel: 0
Ambient temperature
sensitivity:
1
Automatic water: Off
Brightness: 30 %