60 53 362 D3472
D3472.201.10.20.02 07.2016
117
Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.3.5.2
Start cycle on the FLEX lid DAC UNIVERSAL ADVANCED
All settings of the DAC UNIVERSAL were preset according to
"Regular activities [ 59]".
The DAC UNIVERSAL is switched on and ready for operation.
The maximum number of 6 instruments are ready for reprocessing.
The FLEX lid is at hand.
1. Remove the FLEX lid from the holder on the DAC UNIVERSAL.
2. Place the FLEX lid on the counter or put it in the optional lid holder.
3. Plug each instrument onto the corresponding adapter in the FLEX lid.
See "General placement of instruments DAC UNIVERSAL
ADVANCED [ 99]" for more information on attaching the
instruments. The adapters can be replaced if necessary, see
"Mounting adapters [ 53]".
4. Attach a Class 5/6 chemical indicator in the FLEX indicator holder in
the groove on the center post of the FLEX lid. See "Checking the
reprocessing result with an indicator / PCD test device [ 59]" for
more information on using the chemical indicator / PCD test device.
5. Insert the FLEX lid in the cover holder of the DAC UNIVERSAL in
such a way that the instruments are pointing down.
There are instruments on the FLEX lid; the DAC UNIVERSAL is
still open and ready for a cycle.
WARNING
The instruments may not be wrapped when using the FLEX lid, as
cleaning and disinfection would then fail.
IMPORTANT
Do not start the DAC UNIVERSAL without a lid inserted into the lid
holder.
NOTICE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure there
are no objects in the chamber.
NOTICE
The DAC UNIVERSAL and/or the instruments can be damaged if a lid
is not inserted correctly.