1. Press the "Menu/Enter"-Key to go to menu level 1.
2. Press the "Up" or "Down"-Key until the desired option is selected.
3. Press the "Menu/Enter"-Key to confirm the selection.
4. Press the "Up" or "Down"-Key on every menu level until the desired
option is selected.
5. Press the "Menu/Enter"-Key to confirm the selection.
6. Press the "Clear"-Key to return to a menu level.
"Service needed"
(simultaneously 2 seconds of acoustic alarm is given)
Indicates that two years or 3000 cycles have passed since the first use or last
servicing. Reminds the user that service is recommended, see "Display
messages DAC UNIVERSAL STANDARD [ 211]".
"Error"
xx A message is displayed. The list of all display messages is provided in
"Display messages DAC UNIVERSAL STANDARD [ 211]".
"Sterilization error"/
"Disinfection
error"
Sterilization/Disinfection has failed. This is displayed simultaneously with an
acoustic alarm. The list of all display messages is provided in "Display
messages DAC UNIVERSAL STANDARD [ 211]".
"User Maintenance"
Is shown after 200 cycles or every 4 weeks (for 10 cycles a day). Indicates
that cleaning and servicing are needed. For details, see section "Cleaning
and maintenance DAC UNIVERSAL STANDARD [ 192]".
"Low Water"
For manual filling of the water tank. Distilled water must be added.
"Low Oil"
Displayed when the NitramOil #2 lubrication concentrate is almost empty.
"Bad water quality"
Displayed when water quality is inadequate.
"Remember: User Maintenance"
-
Check & Clean
(simultaneously 2 seconds of acoustic alarm is given)
Displayed after 50 cycles. Indicates that cleaning and servicing are due
according to the Check & Clean schedule.
See "Cleaning and maintenance DAC UNIVERSAL STANDARD [ 192]" for
details.