Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 9.5Drain the water tank of the DAC UNIVERSAL STANDARD
båÖäáëÜ
Drain the chamber
1. Insert the STANDARD lid back into the lid holder of the DAC
UNIVERSAL.
2. Press the "Up"- or "Down"-Key until
"Draincham"
is selected.
3. Press the "Menu/Enter"-Key to confirm the selection.
The unit closes automatically.
The tank is emptied.
The lid opens again automatically after 60 seconds.
9.5.2 Drain the water tank via a hose
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ The lid of the DAC UNIVERSAL is inserted.
1. Switch off the water of your water conditioning system.
2. Open the "Water" quick coupling at the back of the DAC UNIVERSAL
unit by twisting it counterclockwise.
3. Detach the hose.
4. Attach a 6/4-mm hose to the "Water" quick coupling at the back of the
unit and twist it clockwise firmly.
5. Place the end of the hose in a sink or bucket.
The hose for draining the water tank is installed.
6. Insert the STANDARD lid in the cover holder of the DAC
UNIVERSAL.
7. Press the "Menu/Enter"-Key to go to menu level 1.
8. Press the "Up" or "Down"-Key until
"Manual"
is selected.
IMPORTANT
If the water tank and/or chamber are not completely empty, repeat all
steps from the beginning until both are completely empty.
IMPORTANT
If you have installed the DAC UNIVERSAL with a permanent water
connection, please remember to turn the water back on.
IMPORTANT
If there is no permanent water connection, fill the water tank with
demineralized or distilled water (water conductivity < 3.0 µS/cm).
IMPORTANT
Steps 1-3 are required only if there is a permanent water connection.
IMPORTANT
The DAC UNIVERSAL must be placed higher than the sink/bucket to
allow the water to drain out.