8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.2Using the menu and display information DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
8.2
Using the menu and display information DAC
UNIVERSAL STANDARD
8.2.1 Display information DAC UNIVERSAL STANDARD
The display indicates the current status of the system as follows:
Display messages_Sirona_Nitram
System status Description
"Ready 121 °C"
The unit is ready for the sensitive program: 121°C / 250°F,
Sterilization period 15 min.
"Ready 134 °C"
The unit is ready for the standard program: 134°C / 273°F,
Sterilization period 3 min.
"Ready 134 °C 18min"
The unit is ready for the prion program: 134°C / 273°F,
Sterilization period 18 min.
"Selected lid type:"
Choose the inserted lid type using the "Up"- and "Down"-Keys. Press the
"Start"-Key to confirm.
"Press safety ring!"
The „Start“-Key has been pressed. The selected lid is shown and the safety
ring must be pressed in order to start the cycle
"In progress"
The cycle starts, the lid closes.
"Door locked"
Indicates that the lid is firmly closed and reprocessing can start.
"Internal flush"
Available only for the STANDARD and FLEX lids.
Cleans the water and air channels of the instruments
"Lubrication"
Available only for the STANDARD lid.
The transmission channels of the instruments are lubricated.
"Cold rinse"
Fills the chamber with cold water for exterior cleaning, pulse-wash (multicyclic
cleaning process) of the instruments, and oil discharge.
"Hot rinse"
Fills the chamber with warm water for exterior cleaning, pulse-wash
(multicyclic cleaning process) of the instruments, and oil discharge.
"Heating"
The chamber is heated at the sterilization temperature and corresponding
pressure set in the selected program.
"Stabilizing"
The selected sterilization temperature and pressure are maintained.
"Back-flush"
Saturated steam is forced through the instrument channels before and after
sterilization and then extracted through the interior channels of the
instruments.
"Sterilization"
;
"Disinfection"
The time display indicates the remaining time for the current sterilization/
disinfection process.
"Expel"
/Drying
(when set to level 1-3)
Standard setting: 0
Purge instruments with air after sterilization.
"Cycle complete"
The cycle is ended. The lid opens slightly.
"Press C to unload"
Indicates that the procedure is finished. Press the "Clear"-Key to open the
DAC UNIVERSAL completely so the instruments can be removed.
"Start wrapped cycle?Instr. treat in
std.cycle before wrap? Press "M" to
confirm"
Indicates that the "Wrapped"-Key has been pressed and reminds the user
that the instruments must have been reprocessed in a standard cycle before
wrapping and use of the WRAPPED lid.