Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.3.2 Reprocessing with the STANDARD lid DAC UNIVERSAL
ADVANCED
The STANDARD lid is used for reprocessing unwrapped rotating
handpieces.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second STANDARD lid.
5.3.2.1
General placement of instruments DAC UNIVERSAL ADVANCED
WARNING
If instruments are attached to an unsuitable adapter or if instruments are
damaged, sterilization and cleaning may fail. In this case, reprocessing
of the instruments must be repeated.
NOTICE
All instruments must be placed straight onto the respective adapter to
prevent damage to the O-rings.
NOTICE
The instruments must be easy to attach to the adapters. If the instrument
does not attach to the adapter easily, it can be damaged. The
manufacturer does not accept responsibility for instruments that are
damaged due to being subjected to force.
NOTICE
Make sure that the maximum permissible height of the instruments is not
exceeded, see "Mounting adapters [ → 53]".
NOTICE
Instruments are damaged by inadequate lubrication.
IMPORTANT
If the DAC UNIVERSAL is started with fewer than 6 instruments placed
in the lid, the lubrication function is not deactivated at the empty
positions. To prevent oily instruments, we recommend placing
instruments in all 6 positions of the lid.