60 53 362 D3472
152 D3472.201.10.20.02 07.2016
7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S
7.3Reset DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
If the alarm lamp flashes, but no acoustic alarm is given, the system has
reached 3000 cycles or the last servicing was 24 months ago. Please
contact your authorized dealer for servicing the DAC UNIVERSAL. The
DAC UNIVERSAL can still be used until servicing if "only" the alarm lamp
flashes and no other message is displayed.
7.3
Reset DAC UNIVERSAL ADVANCED
If the DAC UNIVERSAL was started without attaching the lid in the lid
holder and if the lid holder pressed the safety ring down and can no longer
be opened, please proceed as follows:
✔ The DAC UNIVERSAL is switched on.
✔ The allen wrench included with the unit is at hand.
1. Put the Allen wrench into the small opening at the back of the unit to
fasten the adapter.
2. Press the Allen wrench into the opening until you hear a slight click.
The lid holder of the DAC UNIVERSAL moves up somewhat.
Error 86 is displayed and an acoustic signal sounds.
3. Press the "Up"-Key until the ready screen reappears.
The lid holder of the DAC UNIVERSAL moves up completely.
The DAC UNIVERSAL ready for operation again.
The oil dispenser does not dispense oil. Check whether there is oil in the NitramOil #2 bottle and whether oil is
visible in the oil dispenser.
Replace the NitramOil 2 bottle if it is empty, see "Adding/replacing the
lubrication concentrate".
If the error occurs repeatedly, please contact your authorized dealer.
The DAC UNIVERSAL does not heat up. If the safety system was activated and the DAC UNIVERSAL does not
open, please reset, see "Reset DAC UNIVERSAL ADVANCED [ → 152]".
If the error occurs repeatedly, please contact your authorized dealer.
Description Corrective action