Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD
båÖäáëÜ
8.3.3 Reprocessing with the BASKET lid DAC UNIVERSAL
STANDARD
The BASKET lid is used for reprocessing unwrapped solid instruments.
8.3.3.1
Start cycle on the BASKET lid DAC UNIVERSAL STANDARD
✔ All settings of the DAC UNIVERSAL were preset according to
"Regular activities [ → 153]".
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ A maximum of 5 solid instruments have been prepared for
reprocessing, see "Pre-treatment of solid instruments".
✔ The BASKET lid is at hand.
✔ A NitraClean cleaning tablet is at hand.
1. Remove the BASKET lid from the holder on the DAC UNIVERSAL.
2. Place the BASKET lid on the counter or put it in the optional lid holder.
WARNING
The instruments may not be wrapped when using the BASKET lid, as
cleaning and sterilization would then fail.
IMPORTANT
Do not start the DAC UNIVERSAL without a lid inserted into the lid
holder.