EasyManua.ls Logo

Sirona DAC UNIVERSAL - Filling the Water Tank; Pumping Water through the System

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
D3472.201.10.20.02 07.2016
163
Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.1Regular activities
båÖäáëÜ
8.1.5 Filling the water tank
If no permanent water connection is selected, the DAC UNIVERSAL
water tank must be filled with water manually.
1 filled water tank corresponds to approx. 2 complete cycles of the
standard program.
Capacity of the water tank: 1.8 liters / 0.40 imperial gallons
Minimum filling of the water tank: 0.6 liters / 0.13 imperial gallons for a
standard program at 134°C / 273°F
1. Open the lid of the water tank.
2. Fill demineralized/distilled water (water conductivity < 3.0 µS/cm)
through the sieve into the water tank (min. 0.6 liter, max. 1.8 liter).
8.1.6 Pumping water through the system
Before the DAC UNIVERSAL can be started, you must check whether
water flows into the chamber.
See "Checking water supply to the lids of the DAC UNIVERSAL
STANDARD [ → 194]" for the procedure.
IMPORTANT
Remember to empty the drain tank when it is full, see „Remove and
clean the drain tank [ → 131]".
NOTICE
Pour the water into the tank carefully, otherwise the water can spill into
the interior of the unit next to the tank where it can cause a short circuit.

Table of Contents