Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 9.10Cleaning the empty chamber of the DAC UNIVERSAL STANDARD
båÖäáëÜ
9.10
Cleaning the empty chamber of the DAC
UNIVERSAL STANDARD
Frequency: weekly or every 50 cycles (whichever comes first).
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ The STANDARD lid of the DAC UNIVERSAL is removed and there is
an unobstructed view into the chamber.
✔ There is no water or oil in the chamber or the hoses.
1. Clean the empty chamber with a lint-free cloth and a solution of 70%
isopropyl alcohol alcohol.
2. Insert the STANDARD lid in the cover holder of the DAC
UNIVERSAL.
3. Place a NitraClean tablet in the empty chamber. See "Functions with
the BASKET lid [ → 22]" for more information on using a NitraClean
tablet.
4. Press the "Menu/Enter"-Key twice and select the standard program
using the arrow keys.
5. Confirm your input by pressing the "Menu/Enter"-Key again.
The display shows:
"Ready 134 °C"
.
IMPORTANT
The chamber may not be cleaned with agents containing chlorine.
NOTICE
No instruments may be on the STANDARD lid.
NOTICE
Make sure that the lid is inserted all the way into the holder.
NOTICE
Make sure there are no objects in the chamber.