Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 7.2Errors without a message
båÖäáëÜ
7.2
Errors without a message
Description Corrective action
Leakage. Wipe away any water that may have been spilled after filling.
If the unit continues to leak, please contact your authorized dealer.
There is water in the chamber after end
of cycle
Clean or change the drain filter, see "Cleaning the drain filter [ → 130]" or
"Changing the drain filter [ → 138]".
Check whether the drain hose is blocked or damaged.
Empty the drain tank if it is full.
If the error occurs repeatedly, please contact your authorized dealer.
Condensation in the instruments. If the instruments are to be stored for an extended period, it is important
to purge them (horizontally) with compressed air to remove condensation
from the instruments. Condensation can lead to rust.
IMPORTANT: Use the compressed air at the pressure level (bar)
recommended by the instrument manufacturer.
The instruments must be wrapped after purging.
Adapter is loose. Tighten the adapter with the Allen wrench included.
Wrong liquid. If a liquid other than demineralized or distilled water has been filled into the
water tank, the DAC UNIVERSAL may not be used.
Please contact your authorized dealer.
Sterilization OFF safety ring! (A long
acoustic alarm will be given
simultaneously).
Your DAC UNIVERSAL technician has run a special program.
Disconnect the plug of the DAC UNIVERSAL and reconnect it to the power
supply.
WARNING: Otherwise no sterilization will take place!
The DAC UNIVERSAL does not open. If the lid holder on the safety ring is stuck, please first reset, see "Reset
DAC UNIVERSAL ADVANCED [ → 152]".
IMPORTANT: The autoclave may not be started if the lid is not in the
correct position.
The DAC UNIVERSAL does not take in
any water.
Check whether the automatic water intake is activated in the DAC
UNIVERSAL menu, see "Menu system DAC UNIVERSAL ADVANCED
[ → 77]".
Check whether water from the external water supply is fed to the DAC
UNIVERSAL.
If the error occurs repeatedly, please contact your authorized dealer.
A metallic sound is heard when the pump
is activated.
Please contact your authorized dealer.