Sirona Dental A/S 6Cleaning and maintanance DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 6.13Checking the safety ring of the DAC UNIVERSAL ADVANCED
båÖäáëÜ
1. Clean all available lids with warm water
2. Clean the lids with a solution of 70% isopropyl alcohol.
6.13
Checking the safety ring of the DAC UNIVERSAL
ADVANCED
Frequency: every 4 weeks or every 200 cycles (whichever comes first).
✔ The DAC UNIVERSAL is switched on and ready for operation.
1. Insert the STANDARD lid in the cover holder of the DAC
UNIVERSAL.
If the standard program (134°C/3 min) is already selected, jump
to nr. 5 below.
2. Enter the "program selection“ menu section.
3. Press the “Down”-Key to confirm the selection hereby selecting the
standard program.
4. Return to the ready screen.
5. Make sure the STANDARD lid is selected otherwise use the “Up/
Down”-Keys accordingly on the ready screen, to select the
STANDARD lid.
6. Press the “Start/Stop“-Key to start the cycle.
IMPORTANT
The lid may not be cleaned with agents containing chlorine.
NOTICE
No instruments may be on the STANDARD lid.
NOTICE
Make sure that the lid is inserted all the way into the holder.
NOTICE
Make sure there are no objects in the chamber.
IMPORTANT
If no program is selected, the standard program is used.