EasyManua.ls Logo

Sirona DAC UNIVERSAL - Electromagnetic compatibility

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
44 D3472.201.10.20.02 07.2016
3Product description Sirona Dental A/S
3.6Electromagnetic compatibility Operating Instructions DAC UNIVERSAL
3.6
Electromagnetic compatibility
In order to prevent possible risks caused by electromagnetic interference,
no medical electrical or other equipment should be used near the DAC
UNIVERSAL. The DAC UNIVERSAL complies with the currently valid
standards for electromagnetic compatibility (IEC 60601-1-2).
Only medical devices that comply with IEC 60601-1 may be connected to
the DAC UNIVERSAL. Any non-medical devices connected must also
comply with the IEC 60601-1-1 standard for medical devices.
The DAC UNIVERSAL does not cause any harmful interference; it can be
impaired by electromagnetic interference emitted by other devices, which
can lead to malfunctions.
The DAC UNIVERSAL is referred to as "UNIT" below.
3.6.1 Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in such
an environment.
Emission measurement Conformity Electromagnetic environment - guidelines
RF emissions according to CISPR 11 Group 1 The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions according to CISPR 11 Class B The UNIT is intended for use in all facilities,
including residential areas and in any facilities
connected directly to a public power supply
providing electricity to buildings used for residential
purposes.
Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations / flicker according
to IEC 61000-3-3
coincides

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals