Operating Instructions DAC UNIVERSAL 6.8Remove and clean the drain tank
båÖäáëÜ
2. Unscrew the drain filter from the filter housing counterclockwise using
the filter key.
3. Clean the drain filter thoroughly under running water.
4. Screw the drain filter onto the filter housing clockwise using the filter
key.
5. Screw the lid of the filter housing back on clockwise with the filter key.
6.8
Remove and clean the drain tank
Frequency:
If using manual water: Weekly or every 50 cycles (whichever comes first)
or when it is full.
If using automatic water: Daily or every 50 cycles (whichever comes first)
or when it is full.
✔ The DAC UNIVERSAL is switched off and cooled down.
✔ The waste water tank is cooled down.
✔ You are wearing your protective gear.
NOTICE
Tighten the drain filter well without using force.
NOTICE
Tighten the filter housing well without using force.
NOTICE
Do not start a cycle if the DAC UNIVERSAL is not connected to the drain
tank or drainage system.