Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 9.7Cleaning the drain filter
båÖäáëÜ
9.7
Cleaning the drain filter
Frequency: weekly or every 50 cycles (whichever comes first).
The DAC UNIVERSAL is switched off and cooled down.
You are wearing your protective gear.
The filter key included with the unit is at hand.
1. Unscrew the lid of the filter housing using the filter key.
2. Unscrew the drain filter from the filter housing counterclockwise using
the filter key.
3. Clean the drain filter thoroughly under running water.
4. Screw the drain filter onto the filter housing clockwise using the filter
key.
5. Screw the lid of the filter housing back on clockwise with the filter key.
IMPORTANT
At approx. 10 cycles per day, the drain filter must be changed every 4
weeks, see "Changing the drain filter [ 209]"
NOTICE
Tighten the drain filter well without using force.
NOTICE
Tighten the filter housing well without using force.