Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.2.1 Display information DAC UNIVERSAL ADVANCED
The display will show the current system status. Below examples of these
status messages are shown. Please notice that these examples are from
running a standard program with the standard lid. If another lid is chosen
the arrow indicating the lid type would point to the selected lid.
The display indicates the current status of the system as follows:
The DAC UNIVERSAL has been switched on and is heating.
The machine is ready to run the sensitive program:
121°C, 15 minutes holding time.
The machine is ready to run the standard program:
134°C, 3 minutes holding time.
The machine is ready to run the prion program: 134°C, 18 minutes
holding time.
Indicates that the WRAPPED lid has been selected for the current
program.
The user shall confirm that this cycle is in fact being run with this lid by
pressing the “Right”-Key. If not the cycle is aborted by pressing the “Up”-
Key.
Instruments being used with the WRAPPED lid must have been sterilized
before wrapping.