EasyManua.ls Logo

Sirona DAC UNIVERSAL - 11 Midwest;Iso 4;5-Hole Fixed Connection Adapter

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
230 D3472.201.10.20.02 07.2016
11Accessories Sirona Dental A/S
11.1Adapters for the STANDARD lid Operating Instructions DAC UNIVERSAL
11.1.1.11
Midwest/ISO 4/5-hole fixed connection adapter
Dealing wi th adapters_Sirona_ Nitram
Follow the general safety information in "General placement of
instruments DAC UNIVERSAL ADVANCED [ 99]" or "General
placement of instruments DAC UNIVERSAL STANDARD [ 172]" when
handling the adapters described here.
Attachment
1. Attach the Midwest/ISO 4/5-hole highspeed handpiece with fixed
connection to the adapter. When attaching, make sure that the
various tube diameters match.
2. Press the handpiece down slightly while simultaneously twisting the
threaded ring counterclockwise.
3. Pull lightly on the highspeed handpiece to ensure that it is correctly
attached to the adapter.
Removal
1. Hold the handpiece firmly and simultaneously twist the threaded ring
clockwise to loosen it.
2. Detach the Midwest/ISO 4/5-hole highspeed handpiece with fixed
connection from the adapter.

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals