Sirona Dental A/S 3Product description
Operating Instructions DAC UNIVERSAL 3.1Functions
båÖäáëÜ
See "NitraClean cleaning tablet [ 241]" for more information on
NitraClean cleaning tablets.
Sterilization
After external cleaning, the instruments are sterilized. Saturated steam is
forced through the instruments and air is extracted from the chamber.
Three different sterilization programs are available:
134°C (273°F) with a sterilization time of 3 minutes
134°C (273°F) with a sterilization time of 18 minutes
121°C (250°F) with a sterilization time of 15 minutes
To select the program, see "Programs and program selection [ 33]".
Functions BASKET lid with ADVANC ED
For instructions on reprocessing solid instruments see "Reprocessing
with the BASKET lid DAC UNIVERSAL ADVANCED [ 104]" for the DAC
UNIVERSAL ADVANCED and „Reprocessing with the BASKET lid DAC
UNIVERSAL STANDARD [ 177]“ for the DAC UNIVERSAL
STANDARD.
3.1.3 Functions with the WRAPPED lid
The WRAPPED lid is also available for sterilizing 3 reprocessed, wrapped
instruments.
A sterile filter is included with the WRAPPED lid and must be installed by
a service technician.