10Display messages DAC UNIVERSAL STANDARD Sirona Dental A/S
10.1Display messages and what they mean DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
10.1
Display messages and what they mean DAC
UNIVERSAL STANDARD
The display messages have the following meanings:
Code
no.
Description Corrective action
60 The chamber temperature and
the theoretical steam
temperature deviate from each
other more than 2K.
Please contact your authorized dealer.
61 The software has not registered
a change in temperature in the
steam generator for 3 minutes.
Please contact your authorized dealer.
62 The temperature required for
building up pressure for
sterilization was not reached
within 7½ minutes.
Increase the level of the "Ambient temperature sensitivity" settings.
If the water level is too low, fill demineralized or distilled water (water
conductivity < 3.0 µS/cm) into the water tank.
If the DAC UNIVERSAL is in a room with a temperature < 16°C / 61°F,
ensure that operating conditions is fullfilled.
Press the "Clear"-Key to reset and then restart the DAC UNIVERSAL.
If the error occurs repeatedly, please contact your authorized dealer.
63 Incorrect setting of the water
level switch.
Please contact your authorized dealer.
64 The safety system was
activated.
Please contact your authorized dealer.
65 The safety system is defective. Please contact your authorized dealer.
66 The temperature was too high
during sterilization.
Please contact your authorized dealer.
67 The sterilization time was less
than 3 minutes.
The instruments must be sterilized again.
If the error occurs repeatedly, please contact your authorized dealer.
68 The pressure is too high during
the pressure test.
Please contact your authorized dealer.
70 The pressure in the chamber
does not drop quickly enough
after the pressure test.
Drain water from the chamber.
Clean or change the drain filter, see "Cleaning the drain filter [ → 130]" or
"Changing the drain filter [ → 209]".
Make sure the drain hose is not blocked, defective, or kinked.
IMPORTANT: The drain hose must be made of a heat-resistant material
such as PTFE.
Empty the drain tank if it is full.
If the error occurs repeatedly, please contact your authorized dealer.
71 The DAC UNIVERSAL is
blocked due to insufficient water
quality in the water tank.
Drain the water tank if necessary.
Fill it with fresh demineralized or distilled water (water conductivity < 3.0
µS/cm).
If the error occurs repeatedly, please contact your authorized dealer.