60 53 362 D3472
194 D3472.201.10.20.02 07.2016
9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S
9.3Checking water supply to the lids of the DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
Water is pumped through the internal hoses.
6. Interrupt this process as soon as water flows into the chamber by
pressing the "Clear"-Key.
7. Press the "Clear"-Key until the normal display reappears.
The water and oil residue that is in the chamber and hoses will need
to be drained and cleaned out, see section "Drain the chamber" in
"Drain the water tank via the menu [ → 196]".
9.3
Checking water supply to the lids of the DAC
UNIVERSAL STANDARD
Frequency: weekly or every 50 cycles (whichever comes first).
Checking water supply_with FLEX lid
In order to make sure that the DAC UNIVERSAL is able to bring water to
the chamber and internal channels of the STANDARD & FLEX lids
correctly, this operation needs to be tested.
✔ The DAC UNIVERSAL is switched on and ready for operation.
✔ The lid of the DAC UNIVERSAL is removed and there is an
unobstructed view into the chamber.
✔ You are wearing your protective gear.
✔ The STANDARD lid, FLEX lid (if owned) and black lid extension piece
for water are at hand.
✔ The STANDARD lid, the FLEX lid and the chamber are cooled down.
1. Attach the black lid extension piece for water to the bottom of the lid.
2. Place the other end of the black lid extension piece for water on the
bottom adapter in the bottom of the chamber.
IMPORTANT
The water supply to the lid needs only to be checked for the STANDARD
and FLEX lids, since these are the only ones with internal water
channels.